유방 영상 검사에서 Screen-Film, CR, DR에 관한 Image Quality 비교 연구 원문보기
(A) COMPARATIVE STUDY ON IMAGE QUALITY OF SCREEN-FILM, CR AND DR FOR BREAST IMAGE TESTS

  • 저자

    홍동희

  • 학위수여기관

    한서대학교 건강증진대학원

  • 학위구분

    국내석사

  • 학과

    방사선학과 방사선학전공

  • 지도교수

  • 발행년도

    2005

  • 총페이지

    vii, 65p.

  • 키워드

    유방영상검사 Screen-Film CR;

  • 언어

    kor

  • 원문 URL

    http://www.riss.kr/link?id=T10065156&outLink=K  

  • 초록

    The present study compares and analyzes the image quality of detectors in each system for breast image tests. The objective of this study is to generate basic data as criteria for quality control of the breast image test system by comparing the scoring method of ACR Phantom to the physical measuring method of MTF used for quality control in consideration of psychological or clinical aspect of assessment. In order to evaluate the image quality using ACR Phantom, scoring values were compared and significance and reliability of all results were analyzed and rated by means of SPSS (version 10.1), a program for statistics. The analysis of reliability attested to Cronbach's α value of 0.7194, ensuring the outcomes of this study as reliable. From comparison and analysis of Screen-Film, CR(50㎛), CR(100㎛) and DR systems conducted to 4 items of fiber, calcification, mass and total scores for image rating of ACR Phantom, following results could be drawn: In terms of fiber, first of all, CR(50㎛) and DR each gained the highest points with 4.2, followed by Screen-Film with 3.9 and CR(100㎛) with 3.2. Pearson Chi-Square test for significance featuring a significant probability of 0.000(P