본문 바로가기
HOME> 논문 > 논문 검색상세

논문 상세정보

의공학회지 = Journal of biomedical engineering research v.17 no.4, 1996년, pp.535 - 544   피인용횟수: 2

이식형 중이 청각보조기를 위한 진동 트랜스듀서의 설계
Design of Vibrating Transducer for Implantable Middle Ear Hearing Aid

박형욱    (경북대학교 전자전기공학부, 경북대학교병원 의공학과, 경북대학교 의과대학 이비인후과학교실  );
  • 초록

    In this paper, we analyzed the coil-magnet type vibrating transducer for the implantable middle ear hearing aid which is appropriate for patient's hearing level, and an experimental transducer system is designed For the objective and quantitative analysis of the transducer, a theoretical equivalent model containing coil, magneto and inner ear is developed To perform effective evaluation of the transducer, a transforming ratio Tr is introduced and its range that is suitable for practical implantable middle ear hearing device is foun4 The result of applying physical parameters of ear system to the proposed analytical model shows that frequency response of the coil magrlet type vibrator is predominantly governed by resistive impedance of the coil rather than inertia effect of the magnet and the inner parameters. In addition, we realized an experimental middle ear hearing aid system to show the theoretical validity of designed system and this will provide the basis of the development for actually implantable system.


  • 주제어

    Implantable middle ear .   Hearing aid .   Vibrating transducer.  

  • 참고문헌 (15)

    1. Electromagnetic ossicular augementation device , J. M. Kartush;M. Tos , Otolaryngologic Clinics of North America / v.28,pp.155-172,
    2. J. Millman;A. Grabel , Microelectronics / v.,pp.387-450;560-561;665-667,
    3. M. P. Norton , Fundamentals of noise and vibration analysis for engineers / v.,pp.238-239,
    4. D. K. Cheng , Field and wave electromagnetics / v.,pp.225-294,
    5. S. A. Talbot , Systems physiology / v.,pp.177-186,
    6. Contactless semi-implantable electromagnetic middle ear device for the treatment of sensorineural hearing loss , A. J. Maniglia;W. H. Ko , Otolaryngologic Clinics of North America / v.28,pp.121-140,
    7. 人工中耳の歷史 , J. I. Suzuki;K. Kodera , Johns. / v.11,pp.507-514,
    8. 補聽器と人工中耳 , T. Tsuiki , Johns. / v.11,pp.533-536,
    9. Partially implantable hearing aid device , A. J. Maniglia , U.S. pat. no.5,015,224 / v.,pp.,
    10. M. C. Pollack , Amplification for the hearing-impaired(3rd edition) / v.,pp.33-38,
    11. L. L. Beranek , Acoustics / v.,pp.70-77,
    12. The history and development of the implantable hearing aid , R. L. Goode;M. L. Rosenbaum;A. J. Maniglia , Otolaryngologic Clinics of North America / v.28,pp.1-16,
    13. W. Welkowitz;S. Deutsch , Biomedical Instruments Theory and Design / v.,pp.260-261,
    14. S. M. Selby , Standard mathematical tables(21st edition) / v.,pp.491-499,
    15. Engineering principles of mechanical stimulation of the middle ear , W. H. Ko;W. L. Zhu;A. J. Maniglia , Otolaryngologic Clinics of North America / v.28,pp.29-41,
  • 이 논문을 인용한 문헌 (2)

    1. 1998. "Proposal of Magnetic Coupling Type Implantable Middle Ear Hearing Aid" 의공학회지 = Journal of biomedical engineering research, 19(5): 487~495     
    2. 1999. "DFMT Electromagentic Transducer for Implantable Middle Ear Hearing Aid" 의공학회지 = Journal of biomedical engineering research, 20(5): 625~632     

 활용도 분석

  • 상세보기

    amChart 영역
  • 원문보기

    amChart 영역

원문보기

무료다운로드
  • NDSL :
유료다운로드
  • 원문이 없습니다.

유료 다운로드의 경우 해당 사이트의 정책에 따라 신규 회원가입, 로그인, 유료 구매 등이 필요할 수 있습니다. 해당 사이트에서 발생하는 귀하의 모든 정보활동은 NDSL의 서비스 정책과 무관합니다.

원문복사신청을 하시면, 일부 해외 인쇄학술지의 경우 외국학술지지원센터(FRIC)에서
무료 원문복사 서비스를 제공합니다.

NDSL에서는 해당 원문을 복사서비스하고 있습니다. 위의 원문복사신청 또는 장바구니 담기를 통하여 원문복사서비스 이용이 가능합니다.

이 논문과 함께 이용한 콘텐츠
이 논문과 함께 출판된 논문 + 더보기