임플란트-지대주-나사의 적합에 관한 연구
A STUDY ON THE FIT OF THE IMPLANT-ABUTMENT-SCREW INTERFACE

김낙형   (조선대학교 치과대학 보철학교실UU0001180  ); 정재헌   (조선대학교 치과대학 보철학교실UU0001180  ); 손미경   (조선대학교 치과대학 보철학교실UU0001180  ); 백대화   (광주전남지방중소기업청CC0187073  );
  • 초록

    Statement of problem : There have been previous studies about considerable variations in machining accuracy and consistency in the implant-abutment-screw interfaces. Purpose : The purpose of this study was to evaluate the machining accuracy and consistency of implant/abutment/screw combinations on two randomly selected implants from each of four manufactures. Material and methods : In this study, screws were respectively used to secure a cemented abutment, to a hexlock implant fixture ; teflon coated titanium alloy screw(Torq-Tite) and titanium alloy screw in Steri--Oss system, gold-plated gold-palladium alloy screw(Gold-Tite) and titanium alloy screw in 3i system gild screw ana titanium screw in AVANA Dental Implant system, and titanium screws in Paragon System. The implants were perpendicularly mounted in polymethyl methacrylate autopolymerizing acrylic resin block(Orthodontic resin, Densply International Inc. USA) by use of dental surveyer. Each abutment screw was secured to the implant with recommended torque value using a digital torque controller. Each screw was again tightened after 10 minutes. All samples were cross sectioned with grinder-polisher unit(Omnilap 2000 SBT Inc) after embeded in liquid unsaturated polyester (Epovia, Cray Valley Inc) Results : There were the largest gaps in the neck areas of screws in hexagonal extension implants which were examined in this study. The leading edge of the abutment screw thread (superior surface) was in contact with the implant body thread, and the majority of the contacting surfaces were localized to the middle portion of the mating threads. Considerable variation in the contacting surfaces was noted in the samples evaluated. Amounts of contact in the abutment screw thread were larger for assemblies with Gold-Tite screw, gold alloy screw. Torq-Tite screw than those with titanium screws. The findings of intimate contact between the screw and screw seat were seen in all samples, regardless of manufacturers. However, microgap between the head and lateral neck surface of the screw and the abutment could be dectected in all samples. The findings of intimate contact between the platform of the implant and the bottom of the abutment were consistent in all samples, regardless of manufacturers. However, microgaps between the lateral surface of external hex of the fixture and the abutment could be dectected in all samples. Conclusion : Considerable variations in machining accuracy and consistency were noted in the samples and the implant-abutment-screw interfaces were incomplete. From the results of this study, further development of the system will be required, including improvements in pattern design.


  • 주제어

    Maching accuracy .   Hexlock implant fixture .   Torque-Tite .   Gold-Tite .   Gold screw .   Titanium screw.  

  • 참고문헌 (36)

    1. Zarb HA, Schmitt A. The longitudinal clinical effectiveness of osseointegrated dental implants: The Toronto study. Part III : Problems and complications encounted. J Prosthet Dent 1990;64:185-194 
    2. Binon PP. Evaluation of three slip fit hexagonal implants. Implant Dent 1996;5:253-248 
    3. Binon PP. The effect of implant/abutment hexagonal misfit on screw joint stability. Int J Prosthodont 1996;9:149-152 
    4. Eckert SE, Wollan PC. Restrospective review of 1170 endosseous implants placed in partially edentulous jaws. J Prosthet Dent 1998;79:415-421 
    5. Burguette RL, Johns MR, King TB, Patterson EA. Tightening characteristic for screwed joint in osseointegrated dental implant. J Prosthet Dent 1994;71:592-599 
    6. Carr AB, Brunski JB. Hurley E. The effect of finishing and polishing techniques in implant cylinder preload;9th Annual Meeting of the Academy of Osseointergration (Abstract). Int J Oral Maxilofac Implants 1995;10:123-124 
    7. Besimo CE, Guindy JS, Lewetag D, Meyer J. Prevention of bacterial leakage into and from prefabricated screw-retained crowns on implants in vitro. Int J OralMaxillofac Implants 1999;14:654-660 
    8. Schulte J K. External hex manufacturing tolerances of six implant systems: A pilot study. Implant Dent 1994;3:51-53 
    9. Breeding LC, Dixon DL, Nelson EW, Tietge JD, Torque required to loosen single- tooth implant abutment screw before and after simulated function. Int J Proshodont 1993;6(5): 435-439 
    10. Patyk A, Laugisch J, Steinberg JM. Hohlraumversiegelung bei implantatgetragenen Suprakonstruktionenmit Hilfe Acetatvernetzter Silikone. Z Zahnarz Implantol 1997;13:195-198 
    11. Byren D, Houston F, Cleary R. Claffey N. The fit of cast and prernachined implant abutments. J Prosthet Dent 1998;80:184-192 
    12. Brunski JE, Hipp JE. In the forces on endosteal implant; A measurement system and biomechanical consideration. J Prosthet Dent 1984;51:82 
    13. Dan ET, William RL. Tissue-integrated prostheses complication. Int J Oral Maxillofac Implants 1992;7:477-484 
    14. Han MJ, Chung CH, Choi HC. A study on surface alternation of implant screws after function. J Korean Acad Prosthodont 2001;39:297-305 
    15. Jeong YT, Chung CH, Lee HT. Screw joint stability according to abutment screw materials. J Korean Acad Prosthodont 2001;39:297-305 
    16. Haack JE, Sakaguchi RL, Sun T, Coffey JP. Elongation and preload stress in dental implant abutment screws. Int J Oral Maxillofac Implants 1995;10:529-536 
    17. Jorneus L, Jemt T, Carlsson L. Loads and designs of screw joints for single crowns supported by osseointegrated implants. Int J Oral Maxillofac Implants 1992; 7:353-359 
    18. Martin WC, Woody RD, Miller BH, Miller AW. Implant abutment screw rotations and preloads for four clifferent screw materials and surfaces. J Prosthet Dent 2001 ;86:24-32 
    19. Binon PP. Implants and components :Entering the new millennium. Int J Oral Maxillofac Implants 2000;5:76-94 
    20. Jaarda MJ, Razzoog ME, Gratton DG. Geometric comparison of five interchangeable implant prosthetic retaining screws. J Prosthet Dent 1995;74:373-379 
    21. Quirynen M, Bollen CML, Eyssen H, van Steenberghe D. Microvial penetration along the implant components of the Br${\aa}$ nemark system. An in vitro study. Clin. Oral Implants Res. 1994;5:239-244 
    22. Jemt T, Laney WR, Harris D, Henry, PJ, Krogh P, Polizzi G. Osseointegrated implants for single tooth replacement. A 1-year report from a multicenter prospective study. Int J Oral Maxillofac Implants 1991;6:29-36 
    23. Gross M, Abramovich I, Weiss EI. Microleakage at the abutment-implant interface of osseointegrated implants. Int J Oral Maxillofac Implants 1999;14:94-100 
    24. Boggan RS, Strong TA, Misch CE, Bidez MW. Influence of hex geometry and prostetic table width on statig and fatigue strength of dental implants. J Prosthet Dent 1999;82:436-440 
    25. Sakaguchi RL, Borgersen SE. Nonlinear finite element contact analysis of preload dental implant components. Int J Oral Maxilofac Implant. 1993;655-661 
    26. Binon PP, McHugh MJ. The effect of eliminating implant/ abutment rotational misfit on screw joint stability. Int J Prosthodont 1996;9:511-519 
    27. Carr AB, Brunski JB. Labishak J, Bagley B. Preload comparison between as received and cast to implant cylinder. J Dent Res(IADR Abstract) No.695,1994 
    28. Balshi T. The role of screws in implant systems. Topic V. Int J Oral Maxillofac Implants 1994;9(Spec Suppl):48-63 
    29. Carr AB, Brunski JB, Hurley E. Effect of fabrication, finishing, and polishing procedures on preload in prostheses using conventional gold and plastic cylinders. Int J Oral Maxillofac Implants 1996;11:589-598 
    30. Schulte JK, Coffey J. Comparison of screw retention of nine abutment systems: A pilot study. Implnat Dent 1997;6:28-31 
    31. Sones AD. Complication with osseointegrated implants. J Prosthet Dent 1989 ;62:581-585 
    32. Dixon DL. Breeding LC. Sadler JP, Mckay ML. Comparison of screw loosening, rotation. and deflection among three implant designs. J Prosthet Dent 1995;74:270-8 
    33. Binon PP, Sutter F, Brunski J, Gulbransen H. Weiner R. The role of screws in implant systems. Int J Oral Maxillofac Implants 1994; 9: 48-62 
    34. Lee HT, Kim NH, Chung CH, A study on the stability of implant screw by use of the sealer. J Korean Acad Prosthodont 2001;39:366-375 
    35. Abrahamsson I, Berglundh T, Lindhe J. Soft tissue response to plaque formation at different implant systems. A comparative study in the dog. Clin Oral Implants Res 1998;9:73-79 
    36. Binon PP. Evaluation of matching accuracy and consistency of selected implant, standard abutments and laboratory analog. Int J Prosthodont 1995;8:162-178 
  • 이 논문을 인용한 문헌 (2)

    1. 2007. "" 대한치과보철학회지 = The journal of Korean academy of prosthodontics, 45(6): 769~779     
    2. Shim, Deok-Bo ; Kim, Hee-Jung ; Oh, Sang-Ho ; Chung, Chae-Heon 2008. "Fit of Fixture/Abutment/Screw Interface of Internal Connection Implant Systems" 대한치과턱관절기능교합학회지 = Journal of Korean academy of stomatognathic function and occlusion, 24(3): 283~298     

 저자의 다른 논문

  • 김낙형 (1)

    1. 2001 "Sealer의 사용이 임프란트 나사의 안정성에 미치는 영향" 대한치과보철학회지 = The journal of Korean academy of prosthodontics 39 (4): 366~375    
  • 정재헌 (112)

  • 손미경 (39)

 활용도 분석

  • 상세보기

    amChart 영역
  • 원문보기

    amChart 영역

원문보기

무료다운로드
유료다운로드
  • 원문이 없습니다.

유료 다운로드의 경우 해당 사이트의 정책에 따라 신규 회원가입, 로그인, 유료 구매 등이 필요할 수 있습니다. 해당 사이트에서 발생하는 귀하의 모든 정보활동은 NDSL의 서비스 정책과 무관합니다.

원문복사신청을 하시면, 일부 해외 인쇄학술지의 경우 외국학술지지원센터(FRIC)에서
무료 원문복사 서비스를 제공합니다.

NDSL에서는 해당 원문을 복사서비스하고 있습니다. 위의 원문복사신청 또는 장바구니 담기를 통하여 원문복사서비스 이용이 가능합니다.

이 논문과 함께 이용한 콘텐츠
이 논문과 함께 출판된 논문 + 더보기