Biomechanical Changes of the Lumbar Segment after Total Disc Replacement : Charite$^{(R)}$, Prodisc$^{(R)}$ and Maverick$^{(R)}$ Using Finite Element Model Study

Kim, Ki-Tack    (Department of Orthopedic Surgery, School of Medicine, Kyung Hee University   ); Lee, Sang-Hun    (Department of Orthopedic Surgery, School of Medicine, Kyung Hee University   ); Suk, Kyung-Soo    (Department of Orthopedic Surgery, School of Medicine, Kyung Hee University   ); Lee, Jung-Hee    (Department of Orthopedic Surgery, School of Medicine, Kyung Hee University   ); Jeong, Bi-O    (Department of Orthopedic Surgery, School of Medicine, Kyung Hee University  );
  • 초록

    Objective : The purpose of this study was to analyze the biomechanical effects of three different constrained types of an artificial disc on the implanted and adjacent segments in the lumbar spine using a finite element model (FEM). Methods : The created intact model was validated by comparing the flexion-extension response without pre-load with the corresponding results obtained from the published experimental studies. The validated intact lumbar model was tested after implantation of three artificial discs at L4-5. Each implanted model was subjected to a combination of 400 N follower load and 5 Nm of flexion/extension moments. ABAQUS $^{TM}$ version 6.5 (ABAQUS Inc., Providence, RI, USA) and FEMAP version 8.20 (Electronic Data Systems Corp., Plano, TX, USA) were used for meshing and analysis of geometry of the intact and implanted models. Results : Under the flexion load, the intersegmental rotation angles of all the implanted models were similar to that of the intact model, but under the extension load, the values were greater than that of the intact model. The facet contact loads of three implanted models were greater than the loads observed with the intact model. Conclusion : Under the flexion load, three types of the implanted model at the L4-5 level showed the intersegmental rotation angle similar to the one measured with the intact model. Under the extension load, all of the artificial disc implanted models demonstrated an increased extension rotational angle at the operated level (L4-5), resulting in an increase under the facet contact load when compared with the adjacent segments. The increased facet load may lead to facet degeneration.


  • 주제어

    Artificial disc .   Lumbar spine .   Finite element model .   Rotation angle .   Facet contact load .   von-Mises stress.  

  • 참고문헌 (27)

    1. Ahn MW, Shin KS, Lee JW, Chung IS, Lee CY : Analysis of compression behavior in lumbar spine under simple vertical load vs follower load. J Korean Soc Spine Surg 9 : 78-83, 2002 
    2. Baker DA, Hastings RS, Pruitt L : Study of fatigue resistance of chemical and radiation cross linked medical grade ultrahigh molecular weight polyethylene. J Biomed Mater Res 46 : 573-581, 1999 
    3. Freeman BJ, Davenport J : Total disc replacement in the lumbar spine : a systematic review of the literature. Eur Spine J 15 : S439-S447, 2006 
    4. Chen CS, Feng CK, Cheng CK, Tzeng MJ, Liu CL, Chen WJ : Biomechanical analysis of the disc adjacent to posterolateral fusion with laminectomy in lumbar spine. J Spinal Disorders Tech 18 : 58-65, 2005 
    5. Chen WJ, Lai PL, Niu CC, Chen LH, Fu TS, Wong CB : Surgical treatment of adjacent instability after lumbar spine fusion. Spine 26 : E519-E524, 2001 
    6. Chung SK, Kim YE, Wang KC : Biomechanical effect of constraint in lumbar total disc replacement : a study with finite element analysis. Spine (Phila Pa 1976) 34 :1281-1286, 2009 
    7. David T : Revision of a Charite artificial disc 9.5 years in vivo to a new Charite artificial disc : case report and explant analysis. Eur Spine J 14 : 507-511, 2005 
    8. de Kleuver M, Oner FC, Jacobs WC : Total disc replacement for chronic low back pain : background and a systematic review of the literature. Eur Spine J 12 : 108-116, 2003 
    9. Dooris AP, Goel VK, Grosland NM, Gilbertson LG, Wilder DG : Load-sharing between anterior and posterior elements in a lumbar motion segment implanted with an artificial disc. Spine (Phila Pa 1976) 26 : E122-E129, 2001 
    10. Geisler FH, Blumenthal SL, Guyer RD, McAfee PC, Regan JJ, Johnson JP, et al. : Neurological complications of lumbar artificial disc replacement and comparison of clinical results with those related to lumbar arthrodesis in the literature : results of a multicenter, prospective, randomized investigational device exemption study of Charite intervertebral disc. Invited submission from the Joint Section Meeting on Disorders of the Spine and Peripheral Nerves, March 2004. J Neurosurg Spine 1 : 143-154, 2004 
    11. Goto K, Tajima N, Chosa E, Totoribe K, Kubo S, Kuroki H, et al. : Effects of lumbar spinal fusion on the other lumbar intervertebral levels (three-dimensional finite element analysis). J Orthop Sci 8 : 577-584, 2003 
    12. Ha SK : Finite element modeling of multi-level cervical spinal segments (C3-C6) and biomechanical analysis of an elastomer-type prosthetic disc. Med Eng Phys 28 : 534-541, 2006 
    13. Hitchon PW, Eichholz K, Barry C, Rubenbauer P, Ingalhalikar A, Nakamura S, et al. : Biomechanical studies of an artificial disc implant in the human cadaveric spine. J Neurosurg Spine 2 : 339-343, 2005 
    14. Kim YE, Yun SS, Jung SK : Biomechanical analysis of the implanted constrained and unconstrained ICR types of artificial disc using FE model. J of the KSPE 23 : 176-182, 2006     
    15. Le Huec JC, Basso Y, Aunoble S, Friesem T, Bruno MB : Influence of facet and posterior muscle degeneration on clinical results of lumbar total disc replacement : two-year follow-up. J Spinal Disord Tech 18 : 219-223, 2005 
    16. McAfee PC, Cunningham B, Holsapple G, Adams K, Blumenthal S, Guyer RD, et al. : A prospective, randomized, multicenter Food and Drug Administration investigational device exemption study of lumbar total disc replacement with the CHARITE artificial disc versus lumbar fusion : part II : evaluation of radiographic outcomes and correlation of surgical technique accuracy with clinical outcomes. Spine (Phila Pa 1976) 30 : 1576-1583; discussion E388-E390, 2005 
    17. Natarajan RN, Andersson GB : The influence of lumbar disc height and cross-sectional area on the mechanical response of the disc to physiologic loading. Spine (Phila Pa 1976) 24 : 1873-1881, 1999 
    18. Patwardhan AG, Havey RM, Meade KP, Lee B, Dunlap B : A follower load increases the load-carrying capacity of the lumbar spine in compression. Spine (Phila Pa 1976) 24 : 1003-1009, 1999 
    19. Rohlmann A, Neller S, Claes L, Bergmann G, Wilke HJ : Influence of a follower load on intradiscal pressure and intersegmental rotation of the lumbar spine. Spine (Phila Pa 1976) 26 : E557-E561, 2001 
    20. Rohlmann A, Zander T, Bergmann G : Effect of total disc replacement with prodisc on intersegmental rotation of the lumbar spine. Spine (Phila Pa 1976) 30 : 738-743, 2005 
    21. van Ooij A, Oner FC, Verbout AJ : Complications of artificial disc replacement : a report of 27 patients with the SB Charite disc. J Spinal Disord Tech 16 : 369-383, 2003 
    22. Wang JL, Parnianpour M, Shirazi-Adl A, Engin AE : Rate effect on sharing of passive lumbar motion segment under load-controlled sagittal flexion : viscoelastic finite element analysis. Theoretical and Applied Fracture Mechanics 32 : 119-128, 1999 
    23. White AA III, Panjabi MM : Clinical biomechanics of the spine. Philadelphia : Lippincott, 1978, pp81 
    24. Cunningham BW, Gordon JD, Dmitriev AE, Hu N McAfee PC : Biomechanical evaluation of total disc replacement arthroplasty : an in vitro human cadaveric model. Spine (Phila Pa 1976) 28 : S110-S117, 2003 
    25. Kumaresan S, Yoganandan N, Pintar FA, Maiman DJ : Finite element modeling of the cervical spine : role of intervertebral disc under axial and eccentric loads. Med Eng Physics 21 : 689-700, 1999 
    26. Ng HW, Teo EC, Lee VS : Statistical factorial analysis on the material property sensitivity of the mechanical responses of the C4-C6 under compression, anterior and posterior shear. J Biomech 37 : 771-777, 2004 
    27. Yamamoto I, Panjabi MM, Crisco T, Oxland T : Three-dimensional movements of the whole lumbar spine and lumbosacral joint. Spine (Phila Pa 1976) 14 : 1256-1260, 1989 
  • 이 논문을 인용한 문헌 (1)

    1. 2015. "" Journal of Korean Neurosurgical Society = 대한신경외과학회지, 58(1): 43~49     

 저자의 다른 논문

  • 김기택 (8)

    1. 1999 "척추 측만증을 동반한 흉추에 발생한 동맥류성 골낭종 - 증례 보고 -" 대한골관절종양학회지 = The Journal of the Korean Bone and Joint Tumor Society 5 (1): 76~81    
    2. 2000 "전이성 척추 종양의 수술적 치료" 대한골관절종양학회지 = The Journal of the Korean Bone and Joint Tumor Society 6 (1): 1~9    
    3. 2009 "Extensive Spinal Cord Infarction after Surgical Interruption of Thoracolumbar Dural Arteriovenous Fistula Presenting with Subarachnoid Hemorrhage" Journal of Korean Neurosurgical Society = 대한신경외과학회지 46 (1): 60~64    
    4. 2010 "Lumbar Interbody Fusion Outcomes in Degenerative Lumbar Disease : Comparison of Results between Patients Over and Under 65 Years of Age" Journal of Korean Neurosurgical Society = 대한신경외과학회지 48 (5): 412~418    
    5. 2010 "Surgical Experience of Neglected Lower Cervical Spine Fracture in Patient with Ankylosing Spondylitis" Journal of Korean Neurosurgical Society = 대한신경외과학회지 48 (1): 66~69    
  • Lee, Sang-Hun (4)

  • 이정희 (6)

  • 정비오 (10)

 활용도 분석

  • 상세보기

    amChart 영역
  • 원문보기

    amChart 영역

원문보기

무료다운로드
유료다운로드

유료 다운로드의 경우 해당 사이트의 정책에 따라 신규 회원가입, 로그인, 유료 구매 등이 필요할 수 있습니다. 해당 사이트에서 발생하는 귀하의 모든 정보활동은 NDSL의 서비스 정책과 무관합니다.

원문복사신청을 하시면, 일부 해외 인쇄학술지의 경우 외국학술지지원센터(FRIC)에서
무료 원문복사 서비스를 제공합니다.

NDSL에서는 해당 원문을 복사서비스하고 있습니다. 위의 원문복사신청 또는 장바구니 담기를 통하여 원문복사서비스 이용이 가능합니다.

이 논문과 함께 출판된 논문 + 더보기