Performance of handheld electrocardiogram devices to detect atrial fibrillation in a cardiology and geriatric ward setting

Desteghe, Lien (Faculty of Medicine and Life Sciences, Hasselt University, Martelarenlaan 42, 3500 Hasselt, Belgium ) ; Raymaekers, Zina (Faculty of Medicine and Life Sciences, Hasselt University, Martelarenlaan 42, 3500 Hasselt, Belgium ) ; Lutin, Mark (Department of geriatrics, Jessa Hospital, Stadsomvaart 11, 3500 Hasselt, Belgium ) ; Vijgen, Johan (Heart Center Hasselt, Jessa Hospital, Stadsomvaart 11, 3500 Hasselt, Belgium ) ; Dilling-Boer, Dagmara (Heart Center Hasselt, Jessa Hospital, Stadsomvaart 11, 3500 Hasselt, Belgium ) ; Koopman, Pieter (Heart Center Hasselt, Jessa Hospital, Stadsomvaart 11, 3500 Hasselt, Belgium ) ; Schurmans, Joris (Heart Center Hasselt, Jessa Hospital, Stadsomvaart 11, 3500 Hasselt, Belgium ) ; Vanduynhoven, Philippe (Heart Center Hasselt, Jessa Hospital, Stadsomvaart 11, 3500 Hasselt, Belgium ) ; Dendale, Paul (Faculty of Medicine and Life Sciences, Hasselt University, Martelarenlaan 42, 3500 Hasselt, Belgium ) ; Heidbuchel, Hein (Faculty of Medicine and Life Sciences, Hasselt University, Martelarenlaan 42, 3500 Hasselt, Belgium ) ;
  • 초록  

    Aims To determine the usability, accuracy, and cost-effectiveness of two handheld single-lead electrocardiogram (ECG) devices for atrial fibrillation (AF) screening in a hospital population with an increased risk for AF. Methods and results Hospitalized patients ( n = 445) at cardiological or geriatric wards were screened for AF by two handheld ECG devices (MyDiagnostick and AliveCor). The performance of the automated algorithm of each device was evaluated against a full 12-lead or 6-lead ECG recording. All ECGs and monitor tracings were also independently reviewed in a blinded fashion by two electrophysiologists. Time investments by nurses and physicians were tracked and used to estimate cost-effectiveness of different screening strategies. Handheld recordings were not possible in 7 and 21.4% of cardiology and geriatric patients, respectively, because they were not able to hold the devices properly. Even after the exclusion of patients with an implanted device, sensitivity and specificity of the automated algorithms were suboptimal (Cardiology: 81.8 and 94.2%, respectively, for MyDiagnostick; 54.5 and 97.5%, respectively, for AliveCor; Geriatrics: 89.5 and 95.7%, respectively, for MyDiagnostick; 78.9 and 97.9%, respectively, for AliveCor). A scenario based on automated AliveCor evaluation in patients without AF history and without an implanted device proved to be the most cost-effective method, with a provider cost to identify one new AF patient of €193 and €82 at cardiology and geriatrics, respectively. The cost to detect one preventable stroke per year would be €7535 and €1916, respectively (based on average CHA 2 DS 2 -VASc of 3.9 ± 2.0 and 5.0 ± 1.5, respectively). Manual interpretation increases sensitivity, but decreases specificity, doubling the cost per detected patient, but remains cheaper than sole 12-lead ECG screening. Conclusion Using AliveCor or MyDiagnostick handheld recorders requires a structured screening strategy to be effective and cost-effective in a hospital setting. It must exclude patients with implanted devices and known AF, and requires targeted additional 12-lead ECGs to optimize specificity. Under these circumstances, the expenses per diagnosed new AF patient and preventable stroke are reasonable.


  • 주제어

    Electrocardiogram .   Atrial fibrillation .   Screening .   Handheld ECG .   Cost-effectiveness.  

 활용도 분석

  • 상세보기

    amChart 영역
  • 원문보기

    amChart 영역

원문보기

무료다운로드
  • 원문이 없습니다.
유료다운로드

유료 다운로드의 경우 해당 사이트의 정책에 따라 신규 회원가입, 로그인, 유료 구매 등이 필요할 수 있습니다. 해당 사이트에서 발생하는 귀하의 모든 정보활동은 NDSL의 서비스 정책과 무관합니다.

원문복사신청을 하시면, 일부 해외 인쇄학술지의 경우 외국학술지지원센터(FRIC)에서
무료 원문복사 서비스를 제공합니다.

NDSL에서는 해당 원문을 복사서비스하고 있습니다. 위의 원문복사신청 또는 장바구니 담기를 통하여 원문복사서비스 이용이 가능합니다.

이 논문과 함께 출판된 논문 + 더보기