Optimization of an underwater in-situ LaBr3:Ce spectrometer with energy self-calibration and efficiency calibration

Zeng, Z. ; Pan, X. ; Ma, H. ; He, J. ; Cang, J. ; Zeng, M. ; Mi, Y. ; Cheng, J. ;
  • 초록  

    An underwater in-situ gamma-ray spectrometer based on LaBr 3 :Ce was developed and optimized to monitor marine radioactivity. The intrinsic background mainly from 138 La and 227 Ac of LaBr 3 :Ce was well determined by low background measurement and pulse shape discrimination method. A method of self-calibration using three internal contaminant peaks was proposed to eliminate the peak shift during long-term monitoring. With experiments under different temperatures, the method was proved to be helpful for maintaining long-term stability. To monitor the marine radioactivity, the spectrometer's efficiency was calculated via water tank experiment as well as Monte Carlo simulation.


  • 주제어

    Underwater γ-ray spectrometer .   LaBr3(Ce) .   Temperature effect .   Self-calibration .   MC simulation .   Efficiency calibration.  