Clinical biomechanics v.41, 2017년, pp.28 - 33  

Primary stability and viscoelastic displacement of mini-implant system under loading

Lee, J. ; Jeong, Y.H. ; Pittman, J. ; Deguchi, T. ; Johnston, W.M. ; Fields, H.W. ; Kim, D.G. ;
  • 초록  

    Background: The objectives of the current study were to examine the effects of mini-implant diameters on 1) primary stability before bone properties are changed by active peri-implant bone remodeling and 2) the time-dependent displacement of mini-implant systems in bone under a functional radial loading and their associations with bone mineral density. Methods: Twenty one, 8mm length mini-implants (7 each for 1.4mm, 1.6mm, and 2mm diameters) were placed in mandibular sections of human cadavers (4 males and 3 females, average 69.7 (SD 13.1) years of age). Displacement of the mini-implant in wet bone was assessed during initial and subsequent continuous radial loading of 2N in the mediolateral direction for 2h. Mean, standard deviation and coefficient of variation of peri-implant bone mineral density were obtained using histograms of cone-beam computed tomography attenuation values. The cortical thickness along with the miniscrew site was also measured. Findings: The primary stability and displacement of mini-implants in bone were not significantly different between the 3 diameter groups (p>0.147, n=21). Moderate positive correlations of time-dependent viscoelastic displacement (creep) were found with bone mineral density variability independent of the mini-implant diameters (p>0.11). Interpretation: The post-implantation displacements of mini-implant suggested that the orthodontic treatment loading can develop micromotion between the mini-implant and surrounding bone leading to reduction of its primary stability. Current findings also provide an insight that peri-implant bone mineral density variability plays an important role in controlling displacement of the mini-implant, which determine its stability during early and prolonged orthodontic treatment loading periods.


  • 주제어

    Diameter .   Orthodontic loading .   Creep .   Bone mineral density .   Thickness .   Mini-implant.  

 활용도 분석

  • 상세보기

    amChart 영역
  • 원문보기

    amChart 영역

원문보기

무료다운로드
  • 원문이 없습니다.

유료 다운로드의 경우 해당 사이트의 정책에 따라 신규 회원가입, 로그인, 유료 구매 등이 필요할 수 있습니다. 해당 사이트에서 발생하는 귀하의 모든 정보활동은 NDSL의 서비스 정책과 무관합니다.

NDSL에서는 해당 원문을 복사서비스하고 있습니다. 위의 원문복사신청 또는 장바구니 담기를 통하여 원문복사서비스 이용이 가능합니다.

이 논문과 함께 출판된 논문 + 더보기