Applied energy v.212, 2018년, pp.162 - 174   SCI SCIE
본 등재정보는 저널의 등재정보를 참고하여 보여주는 베타서비스로 정확한 논문의 등재여부는 등재기관에 확인하시기 바랍니다.

Density of carbon dioxide with impurities by Coriolis flow meter, oscillation-type densitometer and equations of state

Nazeri, Mahmoud (Research Centre for Carbon Solutions (RCCS), School of Engineering and Physical Sciences, Heriot-Watt University, Edinburgh EH14 4AS, United Kingdom ) ; Maroto-Valer, M. Mercedes (Research Centre for Carbon Solutions (RCCS), School of Engineering and Physical Sciences, Heriot-Watt University, Edinburgh EH14 4AS, United Kingdom ) ; Jukes, Edward (KROHNE Ltd, 34-38 Rutherford Drive, Park Farm Industrial Estate, Wellingborough, Northants NN8 6AE, United Kingdom ) ;
  • 초록  

    Abstract Transporting CO 2 from the capture plant to the storage site may encounter challenges in the flow metering due to the unusual physical properties of CO 2 – especially when containing impurities. Building on the authors’ previous work on the effect of impurities in CO 2 mixtures on the performance of Coriolis flow meters, in this work, the accuracy of the density measurements of CO 2 mixtures was investigated using a Coriolis meter and a high-pressure Anton Paar densitometer. The measured density data were then compared to the Peng-Robinson cubic equation of state with CO 2 volume correction (PR-CO 2 ) as well as to the multi-parameter GERG-2008 equation of state. The Average Absolute Relative Deviations (AARD) of the density predictions with PR-CO 2 and GERG EOSs from the measured densities with the Anton Paar densitometer are 1.2% and 1.0%, respectively. While the Coriolis meter density values were consistently lower than the actual value in the gas phase. A correlation was obtained to correct the measured densities in the gas phase by fitting the measured densities to the prediction values using PR-CO 2 and GERG EoSs at elevated pressures and temperatures. The AARD was reduced from 26.1% to 2.3% using PR-CO 2 and from 25.4% to 3.6% using GERG EoSs. Highlights Flow measurement challenges of CO 2 -rich mixtures in pipelines were investigated. Mass flow and densities of CO 2 with impurities were measured by a Coriolis meter. The densities were compared to densities measured by oscillation-type densitometer. Peng-Robinson-CO 2 and GERG EoSs were evaluated using obtained densities. A correlation for the density of CO 2 -rich fluids by Coriolis meter is proposed.


  • 주제어

    Coriolis flow meter .   Density measurement .   Carbon dioxide (CO2) with impurities .   Densitometer .   Equation of State (EoS) .   CO2 pipeline.  

 활용도 분석

  • 상세보기

    amChart 영역
  • 원문보기

    amChart 영역

원문보기

무료다운로드
  • 원문이 없습니다.

유료 다운로드의 경우 해당 사이트의 정책에 따라 신규 회원가입, 로그인, 유료 구매 등이 필요할 수 있습니다. 해당 사이트에서 발생하는 귀하의 모든 정보활동은 NDSL의 서비스 정책과 무관합니다.

원문복사신청을 하시면, 일부 해외 인쇄학술지의 경우 외국학술지지원센터(FRIC)에서
무료 원문복사 서비스를 제공합니다.

NDSL에서는 해당 원문을 복사서비스하고 있습니다. 위의 원문복사신청 또는 장바구니 담기를 통하여 원문복사서비스 이용이 가능합니다.

이 논문과 함께 출판된 논문 + 더보기